MilliRef — Handling, Storage & Care
Practical instructions for unpacking, handling, storage, cleaning and troubleshooting of the MilliRef Ag/AgCl reference electrode.
Quick Start
Prepare
Remove the storage vessel and rinse the electrode with DI/distilled water.Measure
Place the electrode into the measurement system.After use
Rinse with DI/distilled water and return the tip to 3 M KCl.
Use & Limitations
Designed for aqueous solutions only.
Intended as a reference electrode, not a current-carrying electrode.
Recommended primarily for short-term measurements.
Appropriate electrical shielding is recommended to improve measurement signal and stability.
Storage
Always store the MilliRef tip immersed in 3 M KCl.
Keep the electrode in its storage vessel filled with 3 M KCl.
Keep the sensing tip wet and minimize unnecessary exposure to air.
Store in a cool place protected from direct sunlight.
Troubleshooting
Unstable potential
Rinse the electrode and verify that the tip has not dried out.
Electrode was stored dry
Place the electrode in fresh 3 M KCl and allow it to recondition before testing again.
Measurement still unstable
Compare the response with a known good reference electrode.
Still having problems?
Cleaning
Routine cleaning
Rinse with distilled/deionized water before and after use.
After contaminated samples
Use a rinse solution compatible with the sample and electrode materials, followed by DI water.