MilliRef — Handling, Storage & Care

Practical instructions for unpacking, handling, storage, cleaning and troubleshooting of the MilliRef Ag/AgCl reference electrode.

Quick Start

  1. Prepare
    Remove the storage vessel and rinse the electrode with DI/distilled water.

  2. Measure
    Place the electrode into the measurement system.

  3. After use
    Rinse with DI/distilled water and return the tip to 3 M KCl.

Quick start icon

Use & Limitations

  • Designed for aqueous solutions only.

  • Intended as a reference electrode, not a current-carrying electrode.

  • Recommended primarily for short-term measurements.

  • Appropriate electrical shielding is recommended to improve measurement signal and stability.

  • Not recommended for very low ionic strength or contamination-sensitive electrolytes.

Use and limitations icon

Storage

Always store the MilliRef tip immersed in 3 M KCl.

  • Keep the electrode in its storage vessel filled with 3 M KCl.

  • Keep the sensing tip wet and minimize unnecessary exposure to air.

  • Store in a cool place protected from direct sunlight.

Storage icon

Troubleshooting

Unstable potential

Rinse the electrode and verify that the tip has not dried out.

Electrode was stored dry

Place the electrode in fresh 3 M KCl and allow it to recondition before testing again.

Measurement still unstable

Compare the response with a known good reference electrode.

Still having problems?

→ Contact TrueRef Systems

Troubleshooting icon

Cleaning

Routine cleaning

Rinse with distilled/deionized water before and after use.

After contaminated samples

Use a rinse solution compatible with the sample and electrode materials, followed by DI water.

Cleaning icon